Normal testing
Test | Slice | Time | Settings | Status | Comment | Image slice | Image samples | Start temp | End temp | Video | |||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | 1 | 34:32 | Passes: 4 x 1mm Power: 100% Speed: 0.4 mm/s Samples: 7 | Success | Passes: 1 more than necessary. Collection:
Some pauses mid-procedure because of suction and collector. Collector did not work, but every sample fell beautifully down in the first sample tube | -14 | -14 | ||||||||||||||||||||||||||||||
2 | 2 | 34:32 | Passes: 4 x 1mm Power: 100% Speed: 0.4 mm/s Samples: 7 | Success | Passes: 1 more than necessary. Collection:
Some samples missed the sample tubes. First sample behaved strangely, but was still sampled as intended. | -10 | -12 | ||||||||||||||||||||||||||||||
3 | 3 | 34:32 | Passes: 4 x 1mm Power: 100% Speed: 0.4 mm/s Samples: 7 | Success | Passes: 1 more than necessary. Collection:
| -9 | -11 | ||||||||||||||||||||||||||||||
4 | 4 | 34:32 | Passes: 4 x 1mm Power: 100% Speed: 0.4 mm/s Samples: 7 | Success | Passes: 1 more than necessary. Collection:
Samples hitting sample tube:
| -9 | -9 | ||||||||||||||||||||||||||||||
Frys stod på over natten. Ingen frost på x-ledeskrue | -27 (Area) | -16 (Area) | |||||||||||||||||||||||||||||||||||
5 | 5 | 34:33 | Passes: 4 x 1mm Power: 100% Speed: 0.4 mm/s Samples: 7 | Passes: 1 more than necessary. Collection:
Samples hitting sample tube:
Collector did not rotate after reaching position 2. | -20 | -21 |
Testing resistance
Test | Slice | Time | Settings | Status | Comment | Image slice | Image samples | Start temp | End temp | Video | ||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
1 | 6 | Passes: 4 x 1mm Power: 100% Speed: 0.1 - 1.0 mm/s Samples: 2 x 10 | Success | Spindle had trouble starting. Pausing the program and giving it a push helped, as the spindle doesn't turn off when pausing the program. Passes: 1 more than necessary. Collection:
Samples hitting sample tube:
Speed comments:
Collector did not rotate after reaching position 2. | -15 -9 | -16 -14 | ||||||||||||||||||||||||||||||||||
2 | 7 | Passes: 4 x 1mm Power: 100% Speed: 1.1 - 2.0 mm/s Samples: 2 x 10 |