Application
Possibilities:
- Temperature limit:
- Max heating rate:
- Max cooling rate:
- Size:
- Atmosphere
- ...
Can not be used for:
- ...
Previously used for:
- ...
Tmax = 2650ºC
- Inert or reducing atmosphere controlled by multiple MFC's With flow from 0,1 to 5 L/min
Gas purged into the furnace: Ar (<2000ºC), N2 (1800-2000ºC), He (>2000ºC)
A gas analyzer is connected to the furnace and the concentration of CO in the off-gas will be obtained.
- Real time logging and graphs for temperature and analysis
Safety and operation
Working zone: Ø = 7,5 cm, H = 15 cm
Vacuum-chamber: 0,16 mbar
- Closed graphite chamber for top loading With controlled atmosphere
Sample loading mechanism from bottom of furnace (elevator)
Quenching/cooling chamber (elevator)
Gas safety system
Online gas analysis system, CO
Previously used for:
- Heating quartz and silicon carbide in the range of 25 to 2450ºC in Ar-CO-H2 gas atmosphere.
Insert Picture
Before you begin
- Contact equipment responsible : Delphine Leroy Jonas Einan Gjøvik , leroy@ntnujonas.e.gjovik@ntnu.no
- Equipment may not be used without training.
- Contact room responsible for access : Delphine Leroy Jonas Einan Gjøvik, leroy@ntnujonas.e.gjovik@ntnu.no
Relevant links:
- Use of gas
- Documentation - Graphite tube furnace -ReSiNa furnace - E-214 (imtlab-users restricted)
- Process metallurgy and Casting