Strength testing


TestSliceTimeSettingsStatusCommentImage sliceImage samplesStart tempEnd tempVideo
113:36:43

Passes: 10 x 0.5mm

Power: 100%

Speed: 0.2 - 2.3 mm/s

Samples: 3 x 22

Startheight: 68

Failure

Mill bit broke on sample 2 of 0.4 mm/s


Fracture probably happened during pass transition at 01:24:25.


Seems samples need more spaceing, since tissue is softer. Current spacing is 1 mm.


All samples at 0.3mm/s did not completely detach from the slice.


Collection: 

Dropping

Did not fall

25

 - Review video


Samples hitting sample tube:

Hit 14
Miss 1

2


-16

-14

-18

-16

224:03:06

Passes: 20 x 0.25mm

Power: 100%

Speed: 0.2 - 0.7 mm/s

Samples: 3 x 6

Startheight: 68

Success

Collection: 

Dropping

Did not fall

171



Samples hitting sample tube:

Hit 15
Miss 1

12

-10

-5

-26

-25



332:42:20

Passes: 20 x 0.25mm

Power: 100%

Speed: 0.8 - 2.2 mm/s

Samples: 3 x 15

Startheight: 68

 

Failed

Mill bit broke on sample 1 of 1.9 mm/s


Collection: 

Dropping

Did not fall

276



Samples hitting sample tube:

Hit 115
Miss 1

12

-17

-11

-19

-16


442:09:09

Passes: 20 x 0.25mm

Power: 100%

Speed: 1.7 mm/s

Samples: 45

Startheight: 68

Offset: -6 (same as every test before)

Failed

Mill bit broke on sample 8 on first pass

 - The right side of the slice becomes raised at 20:40 - 20:45 because of too little support from the arms. This caused the first pass of sample 8 to be too deep.


Collection: 

Dropping

Did not fall

25



Samples hitting sample tube:

Hit 11
Miss 1

1

-13

-10

-14

-12


554:20:27

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.1 - 1.7 mm/s

Samples: 8 x 7

Startheight: 68

Offset: -5

Success

Collection: 


Dropping

Did not fall

1.180
1.280
1.380
1.480
1.580
1.671
1.771


The cut quality seems higher on 1.1 mm/s than 1.7. Even though the mill bit did not break i would recommend max speed at 1.3 mm/s.



Samples hitting sample tube:

Hit 126
Miss 1

28

-12

-11

-17

-14


Confirmatory testing


TestSliceTimeSettingsStatusCommentImage sliceImage samplesStart tempEnd tempVideo
6652:59

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.3 mm/s

Samples: 6

Startheight: 68

Offset: -5

Success

Collection: 

Dropping

Did not fall

60


Samples hitting sample tube:

Tube->123456
Hit5, 6




Miss

1, 2, 3, 4






-12

-10



7752:54

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.3 mm/s

Samples: 6

Startheight: 68

Offset: -5

Success

Collection: 

Dropping

Did not fall

60


Samples hitting sample tube:

Tube->1234567
Hit


3

6
Miss

1

2



4

5


-18

-10

-18

-15


8817:18

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.3 mm/s

Samples: 6

Startheight: 68

Offset: -5

Failed

The "pipes" in the lungs are much harder than the rest of the lung, which caused the slice to break. Increased vibrations can be seen on each pass over these areas.


Collection: 

Dropping

Did not fall

00


Samples hitting sample tube:

Tube->765432
Hit





Miss







-14

-8

-14

-10


981:07:14

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 68

Offset: -5

Success

Collection: 

Dropping

Did not fall

50


Samples hitting sample tube:

Position->123456
Hit

3
5
Miss


2


4


6

-11

-7



10918:49

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 68

Offset: -5

Failure

Collection: 

Dropping

Did not fall




Samples hitting sample tube:

Position->123456
Hit





Miss








Improved collecting system with a funnel and more accurate positions.



-15

-12



119

Passes: 24 x 0.25mm

Power: 100%

Speed: 0.2 - 1.0 mm/s

Samples: 2x 7 

Startheight: 66

Offset: -5

Failure

Had to switch to a different type of mill. same specs, but entire mill bit is ~2mm longer

Mill bit broke at second line of 0.2mm/s


Arteriene i lunger blir for vanskelig å forholde seg til, så vi vil heretter prøve å unngå de i framtidig testing.


Collection: 

Dropping

Did not fall




Samples hitting sample tube:

Position->123456
Hit





Miss









-10

-2



121044:37

Passes: 22 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 67

Offset: -5

Partial Success

Collection: 

Dropping

Did not fall


1


had to readjust the passes

Popping does not work on lung tissue, tissue too soft.


Samples hitting sample tube:

Position->123456
Hit





Miss








-21

-13



1310

1:05:34

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 67

Offset: -5

Success

Collection: 

Dropping

Did not fall

06


Spindle did not cut far enough down. Something must be off in the settings. 


Samples hitting sample tube:

Position->123456
Hit





Miss











1411

43:47

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 68

Offset: -5

Failed

Mill bit broke


Collection: 

Dropping

Did not fall




Forgot to compensate for startheight in previous testings, will therefore return to 68.


Samples hitting sample tube:

Position->123456
Hit





Miss








-17

-9

-18

-13


1512

1:05:26

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 68

Offset: -5

Success

Collection: 

Dropping

Did not fall

60


Samples hitting sample tube:

Position->123456
Hit1234
5, 6
Miss







-13

-5

-14

-9


1513


Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 68

Offset: -5

Failure

Ht a vein


Collection: 

Dropping

Did not fall

00


Samples hitting sample tube:

Position->123456
Hit123456
Miss







-10

-2

-14

-8


1614

1:05:32

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.0 mm/s

Samples: 6

Startheight: 68

Offset: -5

Success

Collection: 

Dropping

Did not fall

60


Samples hitting sample tube:

Position->123456
Hit1
2, 3456
Miss







-9

-4

-14

-10