Strength testing


TestSliceTimeSettingsStatusCommentImage sliceImage samplesStart tempEnd tempVideo
113:36:43

Passes: 10 x 0.5mm

Power: 100%

Speed: 0.2 - 2.3 mm/s

Samples: 3 x 22

Startheight: 68

Failure

Mill bit broke on sample 2 of 0.4 mm/s


Fracture probably happened during pass transition at 01:24:25.


Seems samples need more spaceing, since tissue is softer. Current spacing is 1 mm.


All samples at 0.3mm/s did not completely detach from the slice.


Collection: 

Dropping

Did not fall

25

 - Review video


Samples hitting sample tube:

Hit 14
Miss 1

2




-16

-14

-18

-16


224:03:06

Passes: 20 x 0.25mm

Power: 100%

Speed: 0.2 - 0.7 mm/s

Samples: 3 x 6

Startheight: 68

Success

Collection: 

Dropping

Did not fall

171



Samples hitting sample tube:

Hit 15
Miss 1

12



-10

-5

-26

-25



332:42:20

Passes: 20 x 0.25mm

Power: 100%

Speed: 0.8 - 2.2 mm/s

Samples: 3 x 15

Startheight: 68

 

Failed

Mill bit broke on sample 1 of 1.9 mm/s


Collection: 

Dropping

Did not fall

276



Samples hitting sample tube:

Hit 115
Miss 1

12



-17

-11

-19

-16


442:09:09

Passes: 20 x 0.25mm

Power: 100%

Speed: 1.7 mm/s

Samples: 45

Startheight: 68

Offset: -6 (same as every test before)

Failed

Mill bit broke on sample 8 on first pass

 - The right side of the slice becomes raised at 20:40 - 20:45 because of too little support from the arms. This caused the first pass of sample 8 to be too deep.


Collection: 

Dropping

Did not fall

25



Samples hitting sample tube:

Hit 11
Miss 1

1



-13

-10

-14

-12


55

Passes: 24 x 0.25mm

Power: 100%

Speed: 1.1 - 1.7 mm/s

Samples: 8 x 7

Startheight: 68

Offset: -5








Confirmatory testing


TestSliceTimeSettingsStatusCommentImage sliceImage samplesStart tempEnd tempVideo











  • No labels