Application of furnace
Properties:
- Temperature limit: 2000°C
- Manual heating control
- Thermocouple connection available inside furnace
- Crucible size:
- OD Ø80mm (normally)
- Height 160mm, max 200mm
- Atmosphere:
- Argon or other inert gas must be used
- Some CO gas must be assumed (from trace oxygen reacting with crucible)
- Gas purging
- Sampling
- Crucible visible through glass
Can not be used for
- crucibles in other materials than graphite
Previously used for
- melting/preparation of alloys (Al, Si, Fe, ...) and slags
- sampling of molten metal into top-mounted quartz tube
- immersion of lances, tubes, etc
Before you begin
- Contact equipment responsible:
- Equipment may not be used without training
- Contact room responsible for access.
Relevant links:
Known problems and fixes:
- Vacuum valve down on the back of the furnace tends to lock in the wrong position, causing the vacuum to malfunction?
Documentation
Description | File | Author |
---|---|---|
Operating procedure | Operating instructions Blue induction furnace.doc | Egil Krystad |