The D8 Advance is an instrument from 2005. It is a θ-θ diffractometer utilising CuKα radiation and is equipped with electronically controlled divergence slits and a Våntec-1™  SuperSpeed detector. As well as collecting data at extremely rates, the Våntec-1™ detector offers the ability to collect data in a fixed position mode, taking data “snapshots” of up to 12 degrees 2θ. This makes it an ideal instrument for non-ambient diffraction work, with the capability to rapidly collect large amounts of data for the investigation of structure, reactions and kinetics at high temperature.

The diffractometer has a number of hardware configurations, but is primarily dedicated to non ambient data collection using a selection of low temperature, high temperature and high pressure sample stages covering a temperature range of 80-1673K, under inert and reactive atmospheres up to 20 bar pressure.

 

Configurations

Room Temperature
  • 9 position sample changer or single sample spinner
High/Low temperature
  • MRI TC-Wide Range camera
  • Vacuum, inert or reactive gas atmospheres
  • -190-400 °C with LT stage
  • RT-1200 °C with a choice of two radiant heaters
  • RT-1400 °C with Pt-Rh or Ta strip heaters
High Temperature / High Pressure
  • MRI TCP20 camera
  • Vacuum, inert or reactive gas atmospheres
  • 0-20 bar gas pressure
  • RT-1400 °C with Pt-Rh strip heater
  • RT-1200 °C with radiant heater

Photo: Julian Tolchard

 

Before you plan any experiments:

Contact equipment responsible.