  • CuKα radiation
  • 2.5° primary and secondary Soller slits
  • 60-90 position sample changer
  • LynxEye™ SuperSpeed Detector
  • Variable divergence slit: "V6" means that the divergence slit opens automatically such that the illuminated length  on the sample always remains 6 mm (when using fixed divergence slit, the illuminated length on the sample changes from long to short - this is just geometrical).
Well suited for:
  • Fast XRD scans for phase identification
  • Samples that can wait in a queue for hours before being measured (the engineer will load the samples into the instrument every morning and afternoon)
  • Energy-discriminating detector makes it useful also for the fluorescing elements (Fe, Co, Mn)

...