This state-of-the-art instrument was purchased in 2013, and with the possibility of automatic sample changer for up to 90 samples, makes this instrument the main workhorse of the lab.

Instrument specifications
  • CuKα radiation
  • 2.5° primary and secondary Soller slits
  • 60-90 position sample changer
  • LynxEye™ SuperSpeed Detector
Well suited for:
  • Fast XRD scans for phase identification
  • Samples that can wait in a queue for hours before being measured (the engineer will load the samples into the instrument every morning and afternoon)
  • Energy-discriminating detector makes it useful also for the fluorescing elements (Fe, Co, Mn)
Not so good for:
  • Samples that change fast in air (then the D8 Focus may be better to use, due to the booking system)
  • Samples that doesn't fit the standard sample holders
  • Measurements where you want to avoid the Kα2 contributions (then the A-unit is recommended)

...

Photo: Kristin Høydalsvik

Specifications, standard powder diffraction set-up (all days except Wednesdays)
  • Bragg-Brentano geometry

  • θ–θ operating mode

  • Data collection from 5 to 105° 2θ (typically)

Specifications, grazing-incidence diffraction (GIXRD) set-up (Wednesdays, except holiday periods)

Well-suited for studying the structure of polycrystalline films on substrates.

  • Göbel mirror → parallel beam
  • Incident angle fixed at grazing angle (for example 0.5 to 3°)
  • Scanning detector angle (typically 5 to 80 degrees 2θ)

Before you begin

  • Contact equipment responsible: Kristin Høydalsvik, kristin.hoydalsvik@ntnu.no, room K2-122
  • Equipment may not be used without training. 
  • Access to the lab will be given after training.
Relevant links: