Photo: Kristin Høydalsvik Wells
Specifications, standard powder diffraction set-up (all days except Wednesdays)
Bragg-Brentano geometry
θ–θ operating mode
- 2.5° primary and secondary Soller slits
Data collection from 5 to 105° 2θ (typically)
Variable divergence slit: "V6" means that the divergence slit (the slit between the X-ray source and the sample) opens automatically such that the illuminated length on the sample always remains 6 mm).
Standard measurements:
- Highly crystalline samples: step size: ca. 0.010 °/step
- Crystalline samples: step size: ca. 0.013 °/step
- Low crystallinity samples: step size: ca. 0.044 °/step
(for exact numbers, please consult the measurement PC - ask Lab Manager or substitutes for help if needed).
Specifications, grazing-incidence diffraction (GIXRD) set-up (Wednesdays, except holiday periods)
Well-suited for studying the structure of polycrystalline films on substrates.
...