This diffractometer has a number of hardware configurations, but is primarily dedicated to non ambient data collection using a selection of low temperature, high temperature and high pressure sample stages covering a temperature range of 80-160-1673K1400°C, under inert and reactive atmospheres up to 20 bar pressure.

The Våntec-1™ detector offers the ability to collect data in a fixed position mode, taking data “snapshots” of up to 12 degrees 2θ. This makes it an ideal instrument for non-ambient diffraction work, with the capability to rapidly collect large amounts of data for the investigation of structure, reactions and kinetics at high temperature.

...

  • MRI TC-Wide Range camera
  • Vacuum, inert or reactive gas atmospheres
  • -190-400 °C with LT stage (not in use)
  • RT-1200 °C with a choice of two radiant heaters
  • RT-1400 °C with Pt-Rh or Ta strip heaters
High Temperature /

...

Pressure
  • MRI TCP20 camera
  • Vacuum, inert or reactive gas atmospheres
  • 0-20 2 bar gas pressurepressure  (O2, CO2, N2, 5% N2 + 95 %N2, synthetic air, Ar)
  • RT-1400 °C with Pt-Rh strip heater
  • RT-1200 °C with radiant heater

...