Siemens D5005, with quartz primary monochromator and CuKα1 radiation. Scintillation detector. Optimised The “A-Unit” is a D5005 base goniometer set up in θ-2θ mode and optimised for high quality powder data collection. It is fitted with a focusing primary monochromator and 2.5° Soller slits, giving well collimated monchromatic radiation. It is thus especially suitable for the collection

Well suited for:
  • Collection of data for structure solution and/or Rietveld refinement

...

  • Investigation of strain and crystallite size effects (due to tight and symmetrical line-shape characteristics of the instrument).
Not so good for:
  • Samples containing fluorescing elements (Fe, Co, Mn), depending on the concentration.

 

Image Added

Photo: Julian Tolchard, it is also well suited for the investigation of strain and crystallite size effects.

Instrument specifications:
  • Bragg-Brentano geometry
  • θ–2θ operating mode
  • Data collection from 0.5 5° (but 15°-20° in normal mode) to 140° 2θQuartz primary monochromator
  • Cu Kα1 radiation
  • 2.5° primary and secondary Soller slitsScintillation detector
  • Rotating single sample holder with user selectable rotation speeds
Well suited for:
  • Collection of data for structure solution and/or Rietveld refinement
Not so good for:
  • Samples containing fluorescing elements (Fe, Co, Mn), depending on the concentration.

 

Image Removed

Photo: Julian Tolchard

...

Before you begin:

Equipment may not be used without training. See XRD-lab info.

 